Experiment 11 Atomic Spectra

Purpose:

  1. To observe atomic line spectra from several discharge tube sources.
  2. To measure wavelengths of prominent lines in these spectra with precisions of 0.1% or better
  3. To verify parts of the term diagrams pertaining to the observed lines.

Discussion:

The data which provided the main impetus for the development of quantum mechanics, as well as some of the most important tests of the theory, were atomic spectra. Large devices and long exposure times are necessary for the most precise data. Spectrometers with film strips 30 feet from a diffraction grating and exposure times of several hours are not uncommon. However, quite respectable data can be obtained with a small laboratory spectrometer if care is taken. In this experiment you will use a constant-deviation prism spectrometer to observe spectra from several different sources. The spectrometer gets its name because the prism is ground so that the angle between the incident light beam and the observed beam remains at 90° as the prism is rotated by a dial calibrated in Angstrom units.

Procedure:

Observe the Hydrogen spectrum.

The visible lines are in the Balmer series. Measure as many as you can (four should be observable) and obtain a preliminary set of values of () versus where () is the difference between the dial reading and the known wavelengths of the observed lines. The idea here is that the spectrum is very simple so there should be no ambiguities, and the wavelengths are known exactly from a theoretical expression. Take at least five measurements of each line and use the average for your graph. Notice there is considerable backlash in the dial so it will be necessary to always approach the final setting from the same direction.

Observe the spectra of mercury vapor and sodium vapor.

The lines in the mercury spectrum are commonly used as spectrometer calibration points because of their isolation, prominence and spread. Take five or more readings of each line and find the averages. Use your preliminary calibration curve if necessary to positively identify the lines and construct an improved calibration curve. Include this curve in your report and use it to correct all future measurements. The bright yellow lines in the sodium spectrum are the famous D-lines. Find their wavelength and estimate their spacing. Find the wavelengths of other lines in the sodium spectrum and observe that they are all doublets. Estimate these doublet spacings wherever possible.

Observe the helium spectrum.

It should be possible to obtain wavelengths of from 10 to 15 lines.

Observe spectra from other gaseous discharge tubes available.

Determine the wavelengths of the prominent lines and count the number of weaker lines between the stronger ones. Note sufficient comments to allow completion of column E in the table mentioned below.

Report:

© Department of Physics and Astronomy, University of Iowa, Iowa City, Iowa. All rights reserved.